6
Evaluate

Testing & Data Collection

Duration
45 minutes
Type
Evaluate
Standards
MS-ETS1-3, 7.RP.A.3

Learning Objectives

Students will be able to:

The Testing Protocol

Step 1: Setup (2 min)

  • Place CR box in center of test area
  • Position PM2.5 sensor 3 feet from CR box
  • Make sure fan is OFF
  • Close doors

Step 2: Generate Particles (Teacher Only)

  • Burn incense for 2 minutes
  • Extinguish and remove from room
  • Wait 1 minute for smoke to distribute

Step 3: Baseline Reading (1 min)

  • Record PM2.5 reading - this is "Before"
  • Wait for stable reading (not fluctuating)

Step 4: Run Test (10 min)

  • Turn CR box ON (medium speed)
  • Start timer
  • Do NOT open doors or move around
  • Record PM2.5 every 2 minutes

Step 5: Final Reading

  • At 10 minutes, record final PM2.5 - this is "After"
  • Turn CR box OFF
  • Calculate percent reduction

Understanding PM2.5 Readings

PM2.5 Level Air Quality Color
0-12 μg/m³ Good Green
12.1-35.4 μg/m³ Moderate Yellow
35.5-55.4 μg/m³ Unhealthy for Sensitive Groups Orange
55.5-150.4 μg/m³ Unhealthy Red
150.5+ μg/m³ Very Unhealthy Purple

Note: Sensors may take 1-2 minutes to stabilize. Wait before recording.

Data Collection Sheet

Time (min) PM2.5 (μg/m³) Notes
0 (Before) Baseline - fan OFF
2
4
6
8
10 (After) Final reading

Calculations

Particle Reduction = Before - After = ____ - ____ = ____ μg/m³

Percent Reduction = (Reduction ÷ Before) × 100 = _____%

Fair Testing Principles

Control Variables

Keep these the same for all tests:

  • Room size/location
  • Test duration (10 min)
  • Starting particle concentration
  • Sensor placement (3 ft away)
  • Fan speed setting

Variables We're Testing

Independent: Different CR box designs

Dependent: PM2.5 concentration

Example Calculation

A CR box test shows:

Reduction = 80 - 25 = 55 μg/m³

Percent Reduction = (55 ÷ 80) × 100 = 68.75%

This CR box removed about 69% of particles in 10 minutes!

Troubleshooting

Problem Possible Cause Solution
PM2.5 not decreasing Gaps in seal Check and repair seams
PM2.5 not decreasing Fan direction wrong Verify fan exhausts upward
Readings inconsistent Sensor not stable Wait longer between readings
Baseline too low Not enough particles Generate more smoke
Baseline too high Previous test lingering Ventilate room between tests
← Lesson 5 Lesson 7: Analyzing Results →